C12 465
imc CANSAS Users Manual - Doc. Version 1.9 - 05.12.2014© 2014 imc Meßsysteme GmbH
7.4 C8
Data Sheet Version 1.5 (8 differential analog inputs, individually filtered and conditioned)
Parameter
Value
Remarks
Inputs
8
Measurement modes
DSUB
voltage measurement
thermocouples, RTD
current measurement
with shunt plug ACC/DSUB(M)-I4
Measurement modes
LEMO
voltage measurement
RTD
current measurement
with external shunt
Measurement mode
Thermo-plug type-K (-2T)
thermocouple measurement type-K
Sampling frequency/ Channel
100 Hz
Terminal connection
2x DSUB-15 / 4 channels
or
8x BNC
or
8x 2-pin TK sockets
2x DSUB-9
PHOENIX (MC 1.5/4STF-3.81)
inputs
only for voltage measurement
only for Type K thermocouple
measurement
CAN (in / out), supply (alternatively)
supply
Terminal connection for SL
Inputs
CAN (in / out)
inputs
CAN (in / out)
DC power supply
2 plug DSUB-15/ 4channals
2x DSUB-9
CANSAS-SL-C8-D, (-SUPPLY)
power supply (alternatively)
8x 7-pin LEMO (HGG.1B.307)
2x 10-pin LEMO (HGA.1B.310)
CANSAS-SL-C8-L, (-SUPPLY)
power supply (alternatively)
1x 6-pin LEMO (HGA.1B.306)
for all SL models
Parameter
Value
Remarks
Sampling rate
per channel
Bandbreite
0 Hz to 20 Hz
0 Hz to 10 Hz
-3 dB (voltage measurement)
-3 dB (temperature measurement)
Parameter
Value typ.
min. / max.
Remarks
Isolation:
CAN-Bus
power supply input
analog input
60 V
60 V
no isolation
output to case (CHASSIS)
nominal; testing voltage:300 V(10 s)
nominal; testing voltage:300 V(10 s)
analog reference ground:CHASSIS
Overvoltage protection
250 V
80 V
permanent channel to chassis
<1 ms