EasyManua.ls Logo

IMC CANSAS - Page 466

IMC CANSAS
582 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
466 General Technical Specs
imc CANSAS Users Manual - Doc. Version 1.9 - 05.12.2014© 2014 imc Meßsysteme GmbH
Parameter
typ.
min. / max.
Remarks
Input range
60 V, 20 V, 10 V, 5 V, 2 V, 1 V,
500 mV, 200 mV,
100 mV … 5 mV
Input impedance
1.00 MΩ
492 kΩ
79 kΩ
1%
>135 k
>75 k
differential
±60 V to ±2 V
±1 V to ±50 mV
±20 mV to ±5 mV
Gain uncertainty
0.01%
0.05%
0.02%
0.05%
of reading
±60 V to ±200 mV
±100 mV to ±20 mV
±10 mV to ±5 mV
Gain drift
5 ppm/K T
a
20 ppm/K T
a
T
a
=|T
a
-25°C|; ambient temp. T
a
Offset uncertainty
0.005%
0.005%
0.02%
0.05%
0.01%
0.06%
of input range
60 V to 200 mV
100 mV to 20 mV
10 mV to 5 mV
Offset drift
4 µV/K
0.07 µV/K
< 12 µV/K
< 0.16 µV/K
60 V to 2 V
±1 V to ±5 mV
CMRR common mode rejection
Range
60 V to 2 V
1 V to 5 mV
70 dB
120 dB
> 54 dB
>100 dB
common mode test voltage
50 V
1 V
Noise
51 nV
rms
305 nV
pkpk
range ±5 mV
sampling rate 100 Hz
R
source
= 50 Ω
Bandwidth
0 Hz to 20 Hz
-3 dB

Table of Contents

Other manuals for IMC CANSAS