AM/FM/PM Commands [:SENSe] Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 7-17
[:SENSe]:AFPanalyzer:SPAN <freq>
[:SENSe]:AFPanalyzer:SPAN?
Title: Audio Frequency Span
Description: Sets the audio frequency span in the audio spectrum view. Valid values
are 2 kHz, 5 kHz, 10 kHz, 20 kHz and 70 kHz.
Parameter: <freq>
Default Unit: Hz
Front Panel
Access: Shift-4 (Measure), Audio Spectrum, Span
[:SENSe]:AFPanalyzer:SWEep:TIME <time>
[:SENSe]:AFPanalyzer:SWEep:TIME?
Title: Audio Waveform sweep time
Description: Sets the audio waveform sweep time in the audio waveform view.
Parameter: <time>
Default Unit: Secs
Range: 50 µs to 50 ms.
Front Panel
Access: Shift-4 (Measure), Audio Waveform, Sweep Time
[:SENSe]:AFPanalyzer:SUMMary:DATA?
Title: Summary data
Description: This query returns the measurment values in the Summary view. The
order of the values are as follows:
<demod type> (AM|FM|PM), <RMS Deviation> in % or Hz or Rad
depending on the demod type, <Peak+ Deviation> in % or Hz or Rad
depending on the demod type, <Peak-Dev> in % or Hz or Rad depending
on the demod type,
<Peak-Peak/2 Dev> in % or Hz or Rad depending on the demod type,
<carrier power> in dBm, <carrier freq> in Hz, <Occ BW> in Hz,
<Modulation rate> in Hz, <SINAD> in dB, <THD> in % and
<Distortion> in %. All values are comma separated.
Front Panel
Access: Shift-4 (Measure), Summary.