Commissioning
Type 8098 FLOWave L
9 COMMISSIONING
9.1 Safety instructions
WARNING
Riskofinjuryduetonon-conformingcommissioning.
Non-conforming commissioning could lead to injuries and damage the device and its surroundings.
▶
of the operating instructions.
▶ In particular, observe the safety recommendations and intended use.
▶
9.2 Prerequisites
• The device is installed into the pipe.
• The electrical installation of the device is performed. The device is correctly connected to the functional
earth.
• If your liquid is not water, then make sure that the optional features DF measurement and acoustic trans-
mission factor measurement are activated.
9.3 First commissioning for measuring the ow rate or
for lling containers
Energise the device.
2. Connect the device to the Bürkert Communicator software. Print a pdf report of all the current settings
Bürkert Communicator software from country.burkert.com.
3. On the display module, do the settings of the Quickstart.
4.
If the liquid is not water, then read out the value of the parameter Acoustictransmissionfactor. Refer to
.
•
• liquid might not be measured accurately by
the device.
If the liquid is not water, then read out the value of the parameter DF
•
•
accurately by the device.