75
Commissioning
Type 8098 FLOWave L
 Set the parameter Viscositycompensation for the liquid:
If the liquid is water, then make sure that the parameter Viscositycompensation is set to water. Refer to
.

2
/s, then you can keep the parameter
Viscositycompensation to water.

2
/s, then
set the parameter Viscositycompensation to a value that is adapted to the liquid properties and to the
.
8. Set the parameter Refreshtime to Short.
9. Damping

rate Teach-inbyvolumeow, set the parameter DampingMedium. Refer to
 or .
≥
,  or .
Teach-inbyvolume, set the parameter
DampingNone.
Damping

rate Teach-inbymassowrate, set the parameter DampingMedium. Refer to
 or . Set the parameter Damping of the density to Medium.

,  or . Set the parameter Damping of the density to None. Refer to
.
Teach-inbymass, set the parameter
DampingNone. Set the parameter Damping of the density
to None.
Cut-o function is active and set the
Cut-o or .
Cut-o function is active and set the
Cut-o or .
Density by either using a teach-in procedure, either setting
.
 Set the parameter Kfactor. The K factor applies to both process values volume


of the volume totalizers and mass totalizers and of the pulse outputs are set to Positiveonly and will

set the counting directions to Bothdirections for volume totalizer, refer to
 for mass totalizer.
 Check the correct behaviour of the device by using the menu Simulation.