Starts the run for the first installed plate.
Performs pre-run checks as described in “Processes that the instrument performs during a run” on
page 144.
Scans the barcode for the next plate loaded into the instrument, performs pre-run checks, then
adds the plate to the Run queue. Repeats this step for remaining plates.
IMPORTANT! If you open the dr
awer to add or remove plates, be very careful to remove only plates
that have already been run. If you load a dierent plate in the position of a plate that has not yet been
run, a Link plate error is generated, the instrument pauses, and all injections for the original plate are
cancelled.
After you load the correct original plate in the position, go to the Run Queue screen, select the
cancelled injections, then tap Add to queue.
Pre-run checks and plate alerts in an automated barcode run
In an automated barcode run, pre-run checks are performed when a plate file is link
ed to a plate
position. If pre-run checks do not pass, a
!
(Pre-run checks failed) symbol is displayed at the t
op of
a plate position.
To display the reason for the alert, tap the
!
symbol. A message screen that identifies the failed check
is displayed.
F
or more information, see “Plate alert troubleshooting (automated barcode workflow)” on page 509.
View alert and notification details
1.
If a plate alert symbol (
!
) is displayed for a plate position, tap
!
.
The number of notifications is list
ed with the alert symbol.
Notifications are saved for each user. After you view the notifications, the number of alerts is no
longer displayed. If a dierent user signs in, the number of alerts is displayed until the user views
the notifications.
Chapter 6 S
tart and monitor a run
View alert and notification details
6
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
149