HiDi
Formamide storage and usage
Formamide is an injection solvent that is used to prepare samples. It is not a consumable that is
installed on the instrument as are the other consumables listed in this section. It does not include an
RFID tag on the label.
Cat. No. Description Storage
[1]
4401457 5 mL –25°C to –15°C
If frequent sampling is required, dispense and freeze small aliquots
of HiDi
Formamide. To preserve quality, minimize freeze-thaw
cycles and exposure to air and room temperature. Use of HiDi
Formamide after 8 freeze/thaw cycles has not been validated.
4440753
4 × 5 mL
4311320 25 mL
[1]
S
ee the expiration date on the label. Do not use expired product.
About the SeqStudio
Flex Series Instrument Software v1.0
The SeqStudio
Flex Series Instrument Software v1.0 is the on-instrument, integrated touchscreen
software that runs the SeqStudio
Flex Series Genetic Analyzer.
The software includes the following features:
Complete workflow, from plate file setup to review of primary analysis results
Intuitive display of sample quality, run notifications, and consumables status
4-plate capacity, with interactive plate loading during runs
Ability to prioritize the order of injections and plates
Automated barcode workflow with the internal barcode reader and optional user-supplied external
USB barcode reader
Enhanced results accuracy with spectral autocalibration during every run
Libraries to allow reuse of plate files, injection protocols, and other elements that are required to run
plates
On-instrument learning aids and assistance:
On-instrument tutorial videos with additional videos available on www.thermofisher.com
Demonstration mode that allows use of most features (see “Enable and set up Demo mode” on
page 457)
Step-by-step wizards for instrument maintenance procedures
Smart help and remote session features to request and obtain assistance from Technical
Support (requires connection to Thermo Fisher
Connect)
Optional connection to Thermo Fisher
Connect to access additional features:
Voice commands
Remote plate file creation
Data storage on the cloud
Remote run monitoring from the cloud or a mobile device
Automated cloud analysis with Thermo Fisher Scientific sequencing and fragment analysis
cloud apps
Chapter 1 P
roduct information
About the SeqStudio
Flex Series Instrument Software v1.0
1
28
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide

Table of Contents

Related product manuals