Spatial Calibration maintenance wizard
Spatial Calibration—Time and materials required
Use the Spatial Calibration maintenance wizard to perform a spatial calibration of the capillary array.
This wizard performs the same function described in “Run a spatial calibration” on page 304.
Note: In maintenance wizards that install a capillary array, a spatial calibration is performed as a step in
the wizard.
Item Requirements
Time required ~2–7 minutes
Materials required None
Tap Next to continue.
Spatial Calibration
Tap the option for the spatial calibration.
• Fill and spatial calibration—Use this option ensures that no residual fragments are present in the
capillary array that can interfere with the spatial calibration and signal optimization.
Wait ~7 minutes for the fill array procedure and spatial calibration procedure to complete.
• Spatial calibration—Use this option if you have previously run a spatial calibration and know that
the capillary array is filled with polymer.
Wait ~2 minutes for the spatial calibration procedure to complete.
Spatial Calibration—Results
1.
Tap View details to evaluate the results for the spatial calibration.
For information, see the following sections:
•
“Evaluate the spatial calibration results” on page 306
•
“Export spatial calibration results or report” on page 309
•
“Spatial calibration troubleshooting” on page 522
2.
Tap Finish to exit the wizard.
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
13
412
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide