Manage date and time settings
When you change the date and time setting, you are required to restart the instrument by powering o,
then powering on.
1.
In the home screen, tap Actions4 Settings4Instrument4Date and time.
By default, the time is set to Automatic and the instrument automatically obtains the time from the
network. The Automatic setting is required to use the voice command function. See “Use voice
commands” on page 231.
2.
Press-drag the slider to O if you want to enter the time manually.
3.
Tap Time zone, then select a time zone.
4.
Tap Date/Format, then select the display order for the month, day, and year.
5.
Tap Time/Format, then select a 12-Hour or 24-Hour time display format.
6.
Tap OK.
A message is displayed instructing you to restart the instrument. See “Power o or power on the
instrument” on page 485.
Update the software (administrator only)
All data is retained on the instrument when you update the software.
1.
Select the softwar
e update option in any of the following locations:
About Instrument screen—Tap
Actions4Settings4About4Softwar
e update.
Instrument Settings screen—Tap
Actions4Settings4Instrument4Softwar
e update.
Upcoming Maintenance screen—Tap
(Dashboard) at the t
op right of the home screen.
If the Upcoming Maintenance screen is not displayed, tap
or . Tap P
erform software
update, then tap View info.
The instrument automatically determines if a software update is available on Thermo Fisher
Connect or on a USB drive. If the update is available in both locations, it uses the installation from
the USB drive.
IMPORTANT! Perform a virus scan on a USB drive before inserting it into a port on the
instrument.
Note: It can tak
e up to 15 seconds for the instrument to recognize a USB drive after the drive is
inserted in a USB port. exFAT-formatted USB drives are recommended.