Install or change the capillary array (wizards)
IMPORTANT! Before installing a capillary array, examine the loading-end header to ensure that the
capillary tips are not crushed or damaged.
For the time required, see “Maintenance wizard overview” on page 359.
1.
In the home scr
een, tap
Actions4Maint
enance4Maintenance wizards.
2.
In the Maintenance Wizards screen, tap Change capillary array.
IMPORTANT! Befor
e installing a capillary array, ensure that the capillary tips are not crushed or
damaged. After installing the capillary array, ensure that the fastener on the detection cell heater
door is firmly tightened. If the fastener is not firmly tightened, peaks may be incorrectly detected.
3.
F
ollow the instructions in the wizard to finish the procedure. See “Change Capillary Array
maintenance wizard” on page 382.
IMPORTANT! If an autosampler error or a "power save" message is displayed, or if the CBC is
not in the forward, home position when you need to install the CBC, see “Check the position of the
cathode buer container (CBC)” on page 483.
Chapter 13 Maintain the Instrument
Change and store a capillary array (wizards)
13
374
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide