Export spatial calibration results or report
Note: Spatial and spectral calibration reports show the date of first installation. Install run reports show
the date of the most recent installation.
1.
In the Spatial calibration screen, tap Actions.
2.
Tap the item to export:
Results (ZIP)—Contains the spatial calibration run file (CID), log files, and information that can
be used for troubleshooting.
Results file name format: InstrumentName_SpatialReport_RunDate_Timestamp.zip,
for example: Instrument1_SpatialReport_20201021_121952.zip
Report (PDF)—Contains the results of the spatial calibration for your records.
Report file name format: InstrumentName_SpatialReport_RunDate_Timestamp.pdf, for
example: Instrument1_SpatialReport_20201021_121952.pdf
3.
Select the export location, then tap Next.
For information on connecting to a network drive, see “Connect the software to a network drive
(software settings)” on page 443.
4.
When the export is complete, click Close to close the message box.
Chapter 12 Run calibr
ations and install checks
Run a spatial calibration
12
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide