Overview of install run
Install runs ensure that the instrument meets specifications for sequencing and fragment analysis. For
more information, see “QV20 CRL limits—sequencing install run” on page 346 and “Pass/fail criteria
and results for the fragment install check” on page 350.
You can include multiple injection groups, multiple applications, and/or multiple chemistries on an install
run plate. For example, you can prepare an install run plate that contains replicate injections of the
fragment install standard and both sequencing standards.
Figure 17 Example install run plate with multiple applications
When to perform an install run
You can perform an install run under the following conditions:
If at least one install run for the application type (sequencing or fragment analysis) is not present in
the Install Run History list. This can occur if the install run for the application type has never been
performed, or if all install runs for the application type have been deleted from the instrument.
At any time to verify that the instrument meets specifications.
Chapter 12 Run calibr
ations and install checks
Perform an install run
12
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
329