Screen element Description
Injection controls
Note: Buttons are inactive
if more than 1 injection is
selected or if the action
is not applicable for the
injection.
Inject next—Moves the injection to the next position in the list to be injected. For
information, see “Change the injection list order” on page 163.
Edit & Re-inject—Tap to add re-injections to the run. If needed, change the run
settings. For information, see “Cancel or add injections or specify re-injections
in the Run Queue screen” on page 165. The button is inactive if a spectral
calibration or install run injection is selected or if more than one injection is
selected.
Cancel injection—Stops the current injection or an in-queue injection. The
injection status after cancellation depends on when in the run the injection
was cancelled. See Injection status above. For information, see “Cancel or add
injections or specify re-injections in the Run Queue screen” on page 165.
Add to Queue—Active if you select a cancelled injection. Changes the status to
In queue. The button is inactive if a spectral calibration or install run injection is
selected.
View real-time results in the Run Queue screen
For more information on reviewing results, see:
View real-time results in the home screen” on page 153
“View results for a completed run (Run history)” on page 170
“Use the electropherogram screen” on page 173
Chapter 7, “View results on the instrument”
1.
In the home screen, tap
Run queue.
2.
In the Run Queue scr
een, select an injection or a completed plate with one injection only, then tap
View Plate.
If you select a running or completed injection, the raw or analyzed plot is displayed.
If you select a completed plate with one injection, the plate view is displayed. (The Plate view
button is inactive if you select a plate with more than one injection. Select a single injection to
view more information.)
Chapter 6 S
tart and monitor a run
Use the Run queue to monitor a run on the instrument
6
162
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide