2.
Tap any of the following buttons as needed.
Note: Buttons are active only if you sign in to the software.
•
Ignore & skip—To acknowledge a task without marking it as complete.
•
Mark complete—To mark a task as complete. Marking as complete updates the due date
listed on the screen.
•
Export maintenance history—To generate a CSV file
(InstrumentName_Maintenance_History_Log_Date_Timestamp.csv) that lists all
skipped and completed events.
Daily instrument maintenance tasks
IMPORTANT! Use only the cleaning agents listed in this guide. Use of cleaning agents not listed in this
guide can impair instrument function.
Frequency Task For information, see
Daily Clean the instrument surfaces of dried residue, spilled
buer
, or dirt.
“Clean the instrument” on page 361
Check for leaks and dried residue around the buer-pin
valve, check valve, and array port lock.
If leaks persist, contact Technical Support.
Figure 22 on page 376
“Use Smart Help to request
assistance from Technical Support or
Instrument Service” on page 473
If you use custom run modules that specify an oven
temperature that is below the ambient temperature,
empty the oven condensation reservoir.
“Instrument interior components” on
page 21
Daily or before
each run
Check for bubbles in the pump block and channels.
Use the Remove Bubbles wizard to remove bubbles.
Note: A polymer pouch includes a small reserve volume
that is used for the Remove Bubbles maintenance
wizard, which consumes ~350 μL of polymer. The
reserve volume is sucient to run the wizard ~4
times (including the remove bubbles step during other
maintenance wizards). If you manually run the Remove
Bubbles maintenance wizard >4 times, the volume of
polymer that is available for samples may be depleted.
“Remove bubbles from the polymer
pump (wizard)” on page 377
Visually inspect the capillary tips to ensure that none are
crushed or damaged.
“Instrument interior components” on
page 21
Before each run Check consumables status to see the status for anode
buer container, cathode buer container, polymer, and
capillary array.
“Check consumables status” on
page 66
Chapter 13 Maintain the Instrument
Maintenance schedule
13
356
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide