Radio Frequency Identification
(RFID) technology
■
RFID precautions for use ............................................................. 542
■
Locations of RFID read/write units ..................................................... 543
■
RFID function ....................................................................... 543
■
RFID specifications .................................................................. 544
■
RFID troubleshooting ................................................................ 545
The instrument uses four identical wireless radio frequency identification (RFID) read/write units to
monitor instrument consumables.
RFID precautions for use
WARNING! Radio frequency identification (RFID) could possibly disrupt the operation of patient-
worn and/or implanted active medical devices. To minimize such eects, do not come within 10 cm
of this instrument if you have a patient-worn and/or implanted active medical device.
WARNING! Radio fr
equency identification (RFID) signals from external devices could possibly
disrupt the operation of the instrument read/write units. RFID signals from the instrument RFID
read/write units could possibly disrupt the operation of external RFID devices. To minimize such
eects, do not bring external RFID devices within 10 cm of this instrument during instrument
operation.