EasyManua.ls Logo

Thermo Scientific appliedbiosystems SeqStudio Flex Series - Sequencing and Fragment Applications

Thermo Scientific appliedbiosystems SeqStudio Flex Series
574 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Observation Possible cause Recommended action
"Calibration from manual" is
reported for spectral calibration
status and pull-up peaks are
observed
Auto-spectral calibration was
not used because the sample
possibly contains overlapping
peaks that are labelled with
dierent dyes. For more
information, see “Auto-spectral
calibration during an injection”
on page 150.
IMPORTANT! Spectr
al
calibration status is reported
for informational and tracking
purposes only. This calibration
status does not reflect the
quality of the sample or aect
the results. However, use of
the manual spectral calibration
instead of the auto-spectral
calibration can increase the
likelihood of pull-up peaks.
Perform a manual spectral calibration, then
r
epeat the run. See “Run a spectral
calibration” on page 310.
IMPORTANT! When running samples that
contain overlapping peaks that ar
e labelled
with dierent dyes, manual spectral calibration
is required after changing the capillary array or
moving the instrument.
Sequencing and fragment applications
Observation Possible cause Recommended action
Spikes are present in raw
and/or analyz
ed fluorescence
data
Trace impurities passed the
detector.
For fragment analysis, increasing the
Minimum peak half width setting in the
analysis settings (under Common settings)
can reduce the identification of spikes as
peaks.
Note: Secondary sequencing and fragment
analysis software can recognize and ignore
spikes.
Repeat the injection if necessary.
Poor resolution, low size
quality, or a low-quality
sequencing result is observed
Details: Results are color-
coded based on quality. For
more information, see “Sample
QC and quality alerts” on
page 172.
The samples degraded over
time while on the instrument.
Thermal breakdown of samples is normal.
Samples are stable for 16–24 hours on the
instrument.
Use HiDi
Formamide to prepare the
samples. Sample stability is optimal in HiDi
Formamide.
Prepare the sample according to the protocol
provided with the kits for sample preparation.
See “Sample preparation guidelines” on
page 73.
IMPORTANT! Do not r
esuspend samples in
water.
Appendix A Troubleshooting
Sample and data troubleshooting
A
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
513

Table of Contents

Related product manuals