Device Rated voltage Maximum power
Rated
frequency
Rated power
Instrument 100–240 ±10% VAC
[1]
10 A (1,000 VA) 50/60 Hz 4.2 A (418 VA)
[1]
I
f the supplied power fluctuates beyond the rated voltage, a power line regulator may be required. High or low voltages can adversely
affect the electronic components of the instrument.
Wired and wireless network options and requirements
Thermo Fisher
™
Connect
connection
Local area network (LAN) intranet
connection
Direct connection
Network
connection
Requirement
Wireless A wireless adapter (also referred to as a dongle) is provided with the instrument. The wireless
connection conforms t
o 802.11 b/g/n wireless standards.
Wired The instrument is factory-configured for IPv4 TCP/IP communication and uses an Ethernet
adapter (100/1,000 Mbps) with an RJ45-type connector for local area network (LAN)
connection.
•
An active, tested network jack must be in place before the scheduled installation date.
•
The assigned IT or network specialist from your organization must be available during the
installation to help connect the instrument to your network.
Direct No network is required.
Appendix C Instrument specifications
Wired and wireless network options and requirements
C
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
535