EasyManua.ls Logo

Thermo Scientific appliedbiosystems SeqStudio Flex Series - Evaluate the Spatial Calibration Results

Thermo Scientific appliedbiosystems SeqStudio Flex Series
574 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Evaluate the spatial calibration results
See also “Spatial calibration troubleshooting” on page 522.
1.
Note the pass/fail status of the calibration at the top of the screen.
A Pass
status is required before you can run a spectral calibration.
1
3
2
1
Pass/fail result
2
Capillary CCD pixel marker
3
Tap for more results
2.
E
valuate the spatial calibration profile. Ensure you see the following:
One sharp peak for each capillary. Small shoulders are acceptable.
One marker (+) at the apex of every peak. No o-apex markers.
An even peak profile (all peaks about the same height).
Chapter 12 Run calibr
ations and install checks
Run a spatial calibration
12
306
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide

Table of Contents

Related product manuals