Note: A polymer pouch includes a small reserve volume that is used for the Remove Bubbles
maintenance wizard, which consumes ~350 μL of polymer. The reserve volume is sucient to run the
wizard ~4 times (including the remove bubbles step during other maintenance wizards). If you manually
run the Remove Bubbles maintenance wizard >4 times, the volume of polymer that is available for
samples may be depleted.
For the details on messages that are displayed, see “Check consumables status” on page 66.
Conditioning reagent storage, usage, and limits
Cat. No.
Description Storage
[1]
4393718 Conditioning Reagent Kit 3500/Flex Series
1 unit (single-use)
2–8°C
After removing from storage, use the pouch within
24 hours.
[1]
S
ee the expiration date on the label. Do not use expired product.
On-instrument supported limits Guidelines
Single use or expiration date See the expiration date on the label.
Capillary array storage and usage
WARNING! SHARP The electrodes on the load-end of the capillary array have small, blunt ends that
can lead to piercing injury.
Cat No. Description Storage
A49104 Capillary array 36‑cm SeqS
tudio
™
8 Flex 15–30°C
A49106 Capillary array 50‑cm SeqStudio
™
8 Flex
A49105 Capillary array 36‑cm SeqStudio
™
24 Flex
A49107 Capillary array 50‑cm SeqStudio
™
24 Flex
On-instrument supported limits Guidelines
160 injections when used with Thermo Fisher Scientific
r
eagents, or expiration date
Capillary arrays have been verified for 160
injections when used with Thermo Fisher
Scientific reagents.
If you remove a capillary array from the instrument
for future use, see “Store a capillary array” on
page 375.
Chapter 1 Product information
Instrument consumables handling, usage limits, and expiration
1
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
27