Operating specifications
Component
Specification
Laser
•
Long-life, single-line 505 nm, solid-state laser excitation
source
•
Laser Output power 20 mW
•
Beam divergence 1.4 mrad
LED
•
Emitting color Natural White
•
Luminous Intensity 30 Cd
Electrophoresis voltage Up to 20 kV
Oven temperature Active temperature control from 18°C–70°C
Environmental requirements
Condition
Acceptable range
Installation site Indoor use only
Electromagnetic
interference
Do not use this device in close proximity to sources of strong electromagnetic radiation (for
example, unshielded intentional RF sources). Strong electromagnetic radiation may interfere
with the proper operation of the device.
This equipment has been designed and tested to CISPR 11 Class A. In a domestic
environment it may cause radio interference. You may need to take measures to mitigate
the interference.
Altitude Safety tested up to 2,000 m (6,500 ft.)
Operating conditions 15–30°C (59–86°F), 20–80% relative humidity (noncondensing)
Room temperature should not fluctuate ±2°C during an instrument run.
Transport and storage
conditions
–30 to +60°C ( –22 to +140°F), 5–95% relative humidity
Transient category Installation categories II
Overvoltage category Installation categories II
Vibration Ensure that the instrument is not adjacent to strong vibration sources, such as a centrifuge,
pump, or compressor. Excessive vibration will aect instrument performance.
Pollution degree II
Install the instrument in an environment that is free of pollutants other than non-conductive
pollutants such as dust particles or wood chips. Typical environments with a Pollution
Degree II rating are laboratories and sales and commercial areas.
Appendix C Instrument specifications
Operating specifications
C
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
533