Start and monitor a run
Load and unload a plate .............................................................. 136
Link a plate file and start a run ........................................................ 140
Unlink a plate file ................................................................... 144
Processes that the instrument performs during a run ..................................... 144
Pre-run check messages and plate alerts during a run ................................... 146
Run plates with the automated barcode workflow ....................................... 147
View alert and notification details ...................................................... 149
Auto-spectral calibration during an injection ............................................ 150
View real-time results in the home screen .............................................. 153
Pause the instrument in the home screen ............................................... 157
Use the Run queue to monitor a run on the instrument ................................... 160
Load and unload a plate
Load a plate into the instrument
You can load a plate into the instrument at any time, except when the instrument is injecting samples
from a plate. The drawer is locked during sample injection.
Plates are run in sequential order (A, B, C, D) unless you change the injection order in Run queue.
Note: In the automated barcode workflow, plates are run in the order in which they are loaded into the
drawer. For information, see “Run plates with the automated barcode workflow” on page 147.
Note: Ensure that you use the correct plate retainer for the instrument. See “Plates bases retainers and
septa” on page 539.
IMPORTANT! Ensur
e that the plate retainer is firmly snapped in place on top of the plate (see “Prepare
the plate assembly” on page 77). If the retainer is not snapped in place, the plate assembly can become
jammed in the drawer.
1.
Ensur
e that the Drawer unlocked status is displayed in the home screen.
Positions in the drawer that are empty are marked as Available.
Positions in the drawer that contain a plate display a plate image or are marked as Link a plate
file. In the following figure, all positions are empty and available.
6
136
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide