Size standard normalization feature
For fragment analysis applications, the software includes an optional normalization feature for use
with the GeneScan
600 LIZ
Size Standard v2.0. This feature attenuates signal variations associated
with instrument, capillary array, sample salt load, and injection variability between capillaries and
instruments. Normalization can be applied during data collection.
Based on the settings specified in the run module and the data collected for GeneScan
600 LIZ
Size
Standard v2.0, the software calculates a normalization factor for each sample based on a threshold
setting. The normalization factor is used as a multiplier to adjust the peak height of the sample peaks
relative to the GeneScan
600 LIZ
Size Standard v2.0 peaks.
IMPORTANT! Normalization is not applied t
o samples with failing sizing quality. Select a size standard
definition file appropriate for your application that accurately sizes samples. For example, if your
application includes small fragments that may be obscured by primer peaks, or large fragments that
may not be present due to slower migration rates, specify a size standard definition file that eliminates
these fragments from sizing.
To use the normalization feature:
P
repare each sample with the GeneScan
600 LIZ
Size Standard v2.0.
Use a run module with optimized normalization settings (see the normalization settings as
described in “Run module settings” on page 273).
Select the appropriate normalization size standard in the plate file (see “Size standard definition
files provided” on page 275).
File Name Conventions library
File name convention overview
A file name convention specifies the naming convention for sample data files.
If you do not specify a file name convention for a plate, the following default file name conventions are
used:
Sequencing—Well_Sample_TimeStamp
Fragment analysis—Well_Sample_SampleType_TimeStamp
Chapter 11 Manage libr
ary resources on the instrument
File Name Conventions library
11