(continued)
Condition
Acceptable range
Liquid waste
collection
Dispose of the polymer, buer, reagents and any liquid waste as hazardous waste in
compliance with local and national regulations.
Other conditions Ensure the room is away from any vents that could expel particulate material on the
components.
Avoid placing the instrument and computer adjacent to heaters, cooling ducts, or in direct
sunlight.
WARNING! P
ortable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the instrument,
including cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
WARNING! Use of this equipment adjacent or stack
ed with other equipment should be avoided
because it could result in improper operation.
WARNING! Use of accessories, tr
ansducers, and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
Electrical requirements
CAUTION! Do not unpack or plug in any components until they ar
e configured for the proper
operating voltage by the service representative.
WARNING! F
or safety, the power outlet for the instrument must be accessible at all times. The
instrument must be ~18 cm (7 in.) from a wall or ~36 cm (14 in.) from other instruments or computers.
In case of emergency, you must be able to immediately disconnect the main power supply to all the
equipment. Allow adequate space between the wall and the equipment so that the power cords can
be disconnected in case of emergency.
•
Dedicat
ed line and ground between the instrument and the main electrical service
•
Maximum power dissipation: 1,000 VA, 1,000 W (not including computer or monitor)
•
Mains AC line voltage tolerances must be up to ±10 percent of nominal voltage
Appendix C Instrument specifications
Electrical requirements
C
534
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide