RFID specifications
Table 22 RFID read/write unit specifications
Component Specification
RFID read/write unit
Ultra-Compact Proximal-Type RFID Reader / Writer
Model ASI4000 R2X
Manufactured by ART Finex Co., Ltd.
RF frequency 13.56 MHz
RF output power ~67 mW
RFID tags Texas Instruments RI-I03-112A-03 tags, tested by the manufacturer
to reliably read and write 100,000 times with zero data loss and
retain written data for more than 10 years
Eective range between RFID tag and
internal RFID read/write units
ABC tag: 8.8–9.6 mm
CBC tag: 10.2–10.9 mm
Capillary array tag: 8.3 mm
Polymer pouch tag: 18.4 mm
Typical use range between RFID tag and
internal RFID read/write units
0.5 cm
Minimum separation distance of the
instrument from external RFID read/write
units
10 cm
Minimum separation distance of the
instrument from other wireless technologies
such as WiFi, Bluetooth, or Cellular
92 cm
Minimum separation distance of the
instrument from other laboratory equipment
such as centrifuge or thermal cycler
1 m
Wireless security
RFID tag read/write/re-read with checksum
Password access for use of software
Base-64 encoding of data between the instrument and the
computer
FCC / IC Regulation
Contains FCC ID: 2AM8P-3668
Contains IC: 7805A-3668
Appendix E Radio Frequency Identification (RFID) technology
RFID specifications
E
544
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide