Wash Pump and Channels maintenance wizard
Wash Pump and Channels—Time and materials required
Use the Wash Pump and Channels maintenance wizard to prime the polymer pump with conditioning
reagent, then fill the polymer pump with fresh polymer. This function is automatically performed during
other wizards as needed.
Run this wizard if you observe that polymer has dried in the polymer block or if you suspect
contamination in the pump block.
Item Requirements
Time required ~45 minutes
Materials required
•
Conditioning reagent pouch
•
Empty Anode Buer Container (ABC)
•
Polymer pouch equilibrated to room temperature (15–30°C)
•
Purified water (distilled or deionized)
•
Lint-free lab wipe
Tap Next to continue.
Wash Pump and Channels—Remove polymer pouch, install conditioning reagent
pouch, then wash pump
In this procedure, you replace the polymer pouch with a conditioning reagent pouch, install an empty
anode buer container (ABC) to capture waste, then wash the pump and channels with conditioning
reagent.
1.
Allow the refrigerated polymer to equilibrate to ambient temperature (15–30°C) before use.
2.
Open the instrument doors.
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
13
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
415