Instrument dimensions and clearance
Parameter
Instrument footprint Recommended clearance
Depth ~68 cm (27 inches) ~25 cm (10 inches) at the rear of the
instrument to ensure adequate airflow and
cooling
Width ~70 cm (28 inches) with doors closed
~99 cm (39 inches) with doors open
~18 cm (7 inches) from wall
~36 cm (14 inches) from other instruments
or computers
Height 87 cm (35 inches) ~60 cm (24 inches) from the top of the
instrument
~150 cm (60 inches) from the bench top
Weight ~115 kg (254 lbs)
A bench that is rated for ~150 kg (331 lb) is recommended.