Replenish Buers maintenance wizard
Replenish Buers—Time and materials required
Use the Replenish Buers maintenance wizard to Install a fresh anode buer container (ABC) and
cathode buer container (CBC).
Allow the refrigerated buers to equilibrate to room temperature (15–30°C) before you run the wizard.
Item Requirements
Time required ~4 minutes
Materials required
•
Anode Buer Container (ABC)
•
Cathode Buer Container (CBC)
•
Cathode Buer Container septa
Allow ABC and CBC to equilibrate to room temperature (15–30°C).
Tap Next to continue.
Replenish Buers
WARNING! SHARP The electrodes on the load-end of the capillary array have small, blunt ends that
can lead to piercing injury.
IMPORTANT! If an aut
osampler error or a "power save" message is displayed, or if the CBC is not in
the forward, home position when you need to install the CBC, see “Check the position of the cathode
buer container (CBC)” on page 483.
In this procedure, you install the anode buer container (ABC) and the cathode buer container (CBC).
1.
Allow the r
efrigerated buers to equilibrate to ambient temperature (15–30°C) before use.
2.
Open the instrument doors.
Chapter 13 Maintain the Instrument
Maint
enance wizard procedures
13
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide