Buer storage, usage, and limits
Cat. No.
Description Storage
[1]
4393927 Anode Buer Container 3500/Flex Series (ABC), 4 pack
Contains 1X running buer
2–8°C
[2]
4408256 Cathode Buer Container 3500/Flex Series (CBC), 4 pack
Contains 1X running buer
2–8°C
[2]
[1]
S
ee the expiration date on the label. Do not use expired product.
[2]
Buffers have been qualified to ship at ambient conditions. For a description of the qualification, go to http://find.thermofisher.com/ce/
ambientbuffers/email?CID=fl-we18868.
Instrument
On-instrument supported limits
Lower of:
Guidelines
8-capillary 14 days, 240 injections, or expiration date The buer has been verified for
use for up to 14 days on the
instrument.
24-capillary 14 days, 100 injections, or expiration date
For the details on messages that are displayed, see “Check consumables status” on page 66.
Polymer storage, usage, and limits
Cat. No.
Description Storage
[1]
4393715 POP-4
(384) Performance Optimized Polymer 2–8°C
IMPORTANT! Do not
fr
eeze polymer.
4393710 POP-4
(960) P
erformance Optimized Polymer
4393717 POP-6
(384) Performance Optimized Polymer
4393712 POP-6
(960) Performance Optimized Polymer
4393708 POP-7
(384) Performance Optimized Polymer
4393714 POP-7
(960) Performance Optimized Polymer
[1]
S
ee the expiration date on the label. Do not use expired product.
Pouch size Instrument
On-instrument supported limits
Lower of:
Guidelines
384
samples
8-capillary 14 days, 60 injections, or expiration date The polymer has been verified for use for
up to 14 days on the instrument at the
following temperature ranges:
15–25°C—POP-4
and POP-7
15–30°C—POP-6
At higher temperatures, the limit is 7 days
on the instrument.
24-capillary 14 days, 20 injections, or expiration date
960
samples
8-capillary 14 days, 120 injections, or expiration date
24-capillary 14 days, 50 injections, or expiration date
Chapter 1 Product information
Instrument consumables handling, usage limits, and expiration
1
26
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide