Setting Description
Basecall assignment The basecaller assigns one basecall per position from IUPAC code options:
A, C, G, T, R, Y, S, W, K, M. This setting determines how to analyze a
basecall with a low quality value (QV), where N is any N is any call, per
IUPAC code.
Do not assign Ns to basecalls (default)
Assign Ns to basecalls with QV<X—Basecalls with a QV less than the
X threshold display N instead of the base letter
End base
At PCR stop (default)
After X number of bases
After X number of Ns in X number of bases
After X number of Ns
Note: If you have PCR products with sequences that end while data is still
being collected, select the At PCR stop checkbox.
Note: The default sequencing analysis settings are optimized for PCR
amplicon sequencing. For sequencing of plasmid templates, create analysis
settings with the following selections: for the End base setting, deselect At
PCR stop; for the Mixed base threshold setting, deselect Use Mixed base
identification.
Chapter 11 Manage library resources on the instrument
Analysis settings library
11
292
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide