Change Capillary Array—Remove capillary array
WARNING! SHARP The electrodes on the load-end of the capillary array have small, blunt ends that
can lead to piercing injury.
IMPORTANT! If an aut
osampler error or a "power save" message is displayed, or if the CBC is not in
the forward, home position when you need to remove the capillary array, see “Check the position of the
cathode buer container (CBC)” on page 483.
In this procedure, you fill capillar
y array covers with purified water (if you will store the capillary array)
and remove the capillary array from the instrument.
1.
Prepare for array storage: Fill an array tip cover and array electrode cover with purified water.
Figure 25 Array tip cover and electrode cover for the stored capillary array
1
Array tip cover
2
Electrode cover
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
13
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
383