EasyManua.ls Logo

Thermo Scientific appliedbiosystems SeqStudio Flex Series - Page 405

Thermo Scientific appliedbiosystems SeqStudio Flex Series
574 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.
On the array to be installed, slide the retractor tab on the array to the right until it releases.
Figure 54 Slide the retractor tab to the right
9.
Inser
t the array tip into the array port:
a.
Slide the retractor tab to the left until the array tip reaches the array port lock.
b.
Align the rectangular part of the array tip with the opening in the array port lock.
c.
Insert the array tip into the array port. Press firmly to seat the array tip in the port.
Figure 55 Insert the array tip into the array port
10.
T
urn the array port lock ~1/4 turn clockwise to tighten it. Do not overtighten it.
Chapter 13 Maintain the Instrument
Maint
enance wizard procedures
13
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
405

Table of Contents

Related product manuals