Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 101
*CLS (Clear Status)
(see page 658)
Command Syntax
*CLS
The *CLS common command clears the status data structures, the
device-defined error queue, and the Request- for- OPC flag.
See Also "Introduction to Common (*) Commands" on page 99
"*STB (Read Status Byte)" on page 117
"*ESE (Standard Event Status Enable)" on page 102
"*ESR (Standard Event Status Register)" on page 104
"*SRE (Service Request Enable)" on page 115
":SYSTem:ERRor" on page 394
NOTE
If the *CLS command immediately follows a program message terminator, the output queue
and the MAV (message available) bit are cleared.