Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 207
:CHANnel<n>:VERNier
(see page 658)
Command Syntax
:CHANnel<n>:VERNier <vernier value>
<vernier value> ::= {{1 | ON} | {0 | OFF}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :CHANnel<n>:VERNier command specifies whether the channel's
vernier (fine vertical adjustment) setting is ON (1) or OFF (0).
Query Syntax
:CHANnel<n>:VERNier?
The :CHANnel<n>:VERNier? query returns the current state of the
channel's vernier setting.
Return Format
<vernier value><NL>
<vernier value> ::= {0 | 1}
See Also "Introduction to :CHANnel<n> Commands" on page 190