168 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:ACQuire:DAALias
(see page 658)
Command Syntax
:ACQuire:DAALias <mode>
<mode> ::= {DISable | AUTO}
The :ACQuire:DAALias command sets the disable anti- alias mode of the
oscilloscope.
When set to DISable, anti- alias is always disabled. This is good for cases
where dithered data is not desired.
When set to AUTO, the oscilloscope turns off anti- alias control as needed.
Such cases are when the FFT or differentiate math functions are silent.
The :DIGitize command always turns off the anti- alias control as well.
Query Syntax
:ACQuire:DAALias?
The :ACQuire:DAALias? query returns the oscilloscope's current disable
anti- alias mode setting.
Return Format
<mode><NL>
<mode> ::= {DIS | AUTO}
See Also "Introduction to :ACQuire Commands" on page 163
":ACQuire:AALias" on page 165

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals