Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 381
:SBUS:IIC:ASIZe
(see page 658)
Command Syntax
:SBUS:IIC:ASIZe <size>
<size> ::= {BIT7 | BIT8}
The :SBUS:IIC:ASIZe command determines whether the Read/Write bit is
included as the LSB in the display of the IIC address field of the decode
bus.
Query Syntax
:SBUS:IIC:ASIZe?
The :SBUS:IIC:ASIZe? query returns the current IIC address width setting.
Return Format
<mode><NL>
<mode> ::= {BIT7 | BIT8}
Errors • "- 241, Hardware missing" on page 617
See Also • "Introduction to :SBUS Commands" on page 374
• ":TRIGger:IIC Commands" on page 453
NOTE
This command is only valid on 4 (analog) channel oscilloscope models when the low-speed
IIC and SPI serial decode option (Option LSS) has been licensed.