490 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:UART:BURSt
(see page 658)
Command Syntax
:TRIGger:UART:BURSt <value>
<value> ::= {OFF | 1 to 4096 in NR1 format}
The :TRIGger:UART:BURSt command selects the burst value (Nth frame
after idle period) in the range 1 to 4096 or OFF, for the trigger when in
UART mode.
Query Syntax
:TRIGger:UART:BURSt?
The :TRIGger:UART:BURSt? query returns the current UART trigger burst
value.
Return Format
<value><NL>
<value> ::= {OFF | 1 to 4096 in NR1 format}
See Also "Introduction to :TRIGger Commands" on page 411
":TRIGger:MODE" on page 417
":TRIGger:UART:IDLE" on page 492
":TRIGger:UART:TYPE" on page 498

Table of Contents