EasyManuals Logo

Agilent Technologies InfiniiVision 5000 Series Programmer's Guide

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #450 background imageLoading...
Page #450 background image
450 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:GLITch:QUALifier
(see page 658)
Command Syntax
:TRIGger:GLITch:QUALifier <operator>
<operator> ::= {GREaterthan | LESSthan | RANGe}
This command sets the mode of operation of the glitch pulse width trigger.
The oscilloscope can trigger on a pulse width that is greater than a time
value, less than a time value, or within a range of time values.
Query Syntax
:TRIGger:GLITch:QUALifier?
The :TRIGger:GLITch:QUALifier? query returns the glitch pulse width
qualifier.
Return Format
<operator><NL>
<operator> ::= {GRE | LESS | RANG}
See Also "Introduction to :TRIGger Commands" on page 411
":TRIGger:GLITch:SOURce" on page 452
":TRIGger:MODE" on page 417