384 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:SBUS:SPI:WIDTh
(see page 658)
Command Syntax
:SBUS:SPI:WIDTh <word_width>
<word_width> ::= integer 4-16 in NR1 format
The :SBUS:SPI:WIDTh command determines the number of bits in a word
of data for SPI.
Query Syntax
:SBUS:SPI:WIDTh?
The :SBUS:SPI:WIDTh? query returns the current SPI decode word width.
Return Format
<word_width><NL>
<word_width> ::= integer 4-16 in NR1 format
Errors "- 241, Hardware missing" on page 617
See Also "Introduction to :SBUS Commands" on page 374
":SBUS:MODE" on page 383
":TRIGger:SPI Commands" on page 470
NOTE
This command is only valid on 4 (analog) channel oscilloscope models when the low-speed
IIC and SPI serial decode option (Option LSS) has been licensed.