Status Reporting 9
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 639
Operation Status Condition Register (:OPERegister:CONDition)
The Operation Status Condition Register register hosts these bits:
The :OPERegister:CONDition? query returns the value of the Operation
Status Condition Register.
Name Location Description
RUN bit bit 3 Is set whenever the instrument is not stopped.
WAIT TRIG bit bit 5 Is set by the Trigger Armed Event Register and
indicates that the trigger is armed.
MTE bit bit 9 Comes from the Mask Test Event Registers.
OVLR bit bit 11 Is set whenever a 50Ω input overload occurs.
HWE bit bit 12 Comes from the Hardware Event Registers.