Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 291
:MEASure:RESults
(see page 658)
Query Syntax
:MEASure:RESults?
The :MEASure:RESults? query returns the results of the continuously
displayed measurements. The response to the MEASure:RESults? query is a
list of comma- separated values.
If more than one measurement is running continuously, the
:MEASure:RESults return values are duplicated for each continuous
measurement from the first to last (left to right) result displayed. Each
result returned is separated from the previous result by a comma. There is
a maximum of four continuous measurements that can be continuously
displayed at a time.
When no quick measurements are installed, the :MEASure:RESults? query
returns nothing (empty string). When the count for any of the
measurements is 0, the value of infinity (9.9E+37) is returned for the min,
max, mean, and standard deviation.
Return Format
<result_list><NL>
<result_list> ::= comma-separated list of measurement results
The following shows the order of values received for a single measurement
if :MEASure:STATistics is set to ON.
Measurement label, current, min, max, mean, std dev, and count are only
returned if :MEASure:STATistics is ON.
If :MEASure:STATistics is set to CURRent, MIN, MAX, MEAN, STDDev, or
COUNt only that particular statistic value is returned for each
measurement that is on.
See Also "Introduction to :MEASure Commands" on page 272
":MEASure:STATistics" on page 299
Example Code
' This program shows the InfiniiVision oscilloscopes' measurement
' statistics commands.
' -------------------------------------------------------------------
Option Explicit
Public myMgr As VisaComLib.ResourceManager
Public myScope As VisaComLib.FormattedIO488
Public varQueryResult As Variant
Public strQueryResult As String
Measurem
ent label
current min max mean std dev count

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals