Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 507
:WAVeform:BYTeorder
(see page 658)
Command Syntax
:WAVeform:BYTeorder <value>
<value> ::= {LSBFirst | MSBFirst}
The :WAVeform:BYTeorder command sets the output sequence of the WORD
data. The parameter MSBFirst sets the most significant byte to be
transmitted first. The parameter LSBFirst sets the least significant byte to
be transmitted first. This command affects the transmitting sequence only
when :WAVeform:FORMat WORD is selected. The default setting is
LSBFirst.
Query Syntax
:WAVeform:BYTeorder?
The :WAVeform:BYTeorder query returns the current output sequence.
Return Format
<value><NL>
<value> ::= {LSBF | MSBF}
See Also "Introduction to :WAVeform Commands" on page 502
":WAVeform:DATA" on page 509
":WAVeform:FORMat" on page 511
":WAVeform:PREamble" on page 516
Example Code "Example Code" on page 521
"Example Code" on page 517

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals