EasyManuals Logo

Agilent Technologies InfiniiVision 5000 Series Programmer's Guide

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #570 background imageLoading...
Page #570 background image
570 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
7 Obsolete and Discontinued Commands
:CHANnel<n>:INPut
(see page 658)
Command Syntax
:CHANnel<n>:INPut <impedance>
<impedance> ::= {ONEMeg | FIFTy}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :CHANnel<n>:INPut command selects the input impedance setting for
the specified channel. The legal values for this command are ONEMeg (1
M
) and FIFTy (50).
Query Syntax
:CHANnel<n>:INPut?
The :CHANnel<n>:INPut? query returns the current input impedance
setting for the specified channel.
Return Format
<impedance value><NL>
<impedance value> ::= {ONEM | FIFT}
NOTE
The :CHANnel<n>:INPut command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :CHANnel<n>:IMPedance command (see page 195)
instead.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals