EasyManua.ls Logo

Agilent Technologies InfiniiVision 5000 Series

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 217
:DISPlay:VECTors
(see page 658)
Command Syntax
:DISPlay:VECTors <vectors>
<vectors> ::= {{1 | ON} | {0 | OFF}}
The :DISPlay:VECTors command turns vector display on or off. When
vectors are turned on, the oscilloscope displays lines connecting sampled
data points. When vectors are turned off, only the sampled data is
displayed.
Query Syntax
:DISPlay:VECTors?
The :DISPlay:VECTors? query returns whether vector display is on or off.
Return Format
<vectors><NL>
<vectors> ::= {1 | 0}
See Also "Introduction to :DISPlay Commands" on page 208

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Related product manuals