Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 167
:ACQuire:COUNt
(see page 658)
Command Syntax
:ACQuire:COUNt <count>
<count> ::= integer in NR1 format
In averaging mode, the :ACQuire:COUNt command specifies the number of
values to be averaged for each time bucket before the acquisition is
considered to be complete for that time bucket. When :ACQuire:TYPE is set
to AVERage, the count can be set to any value from 2 to 65536.
Query Syntax
:ACQuire:COUNT?
The :ACQuire:COUNT? query returns the currently selected count value for
averaging mode.
Return Format
<count_argument><NL>
<count_argument> ::= an integer from 2 to 65536 in NR1 format
See Also "Introduction to :ACQuire Commands" on page 163
":ACQuire:TYPE" on page 177
":DIGitize" on page 132
":WAVeform:COUNt" on page 508
NOTE
The :ACQuire:COUNt 1 command has been deprecated. The AVERage acquisition type with
a count of 1 is functionally equivalent to the HRESolution acquisition type; however, you
should select the high-resolution acquisition mode with the :ACQuire:TYPE HRESolution
command instead.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals