186 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:CALibrate:SWITch
(see page 658)
Query Syntax
:CALibrate:SWITch?
The :CALibrate:SWITch? query returns the rear-panel calibration protect
(CAL PROTECT) switch state. The value PROTected indicates calibration is
disabled, and UNPRotected indicates calibration is enabled.
Return Format
<switch><NL>
<switch> ::= {PROT | UNPR}
See Also "Introduction to :CALibrate Commands" on page 179