108 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
*OPC (Operation Complete)
(see page 658)
Command Syntax
*OPC
The *OPC command sets the operation complete bit in the Standard Event
Status Register when all pending device operations have finished.
Query Syntax
*OPC?
The *OPC? query places an ASCII "1" in the output queue when all
pending device operations have completed. The interface hangs until this
query returns.
Return Format
<complete><NL>
<complete> ::= 1
See Also "Introduction to Common (*) Commands" on page 99
"*ESE (Standard Event Status Enable)" on page 102
"*ESR (Standard Event Status Register)" on page 104