EasyManua.ls Logo

Agilent Technologies InfiniiVision 5000 Series

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
108 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
*OPC (Operation Complete)
(see page 658)
Command Syntax
*OPC
The *OPC command sets the operation complete bit in the Standard Event
Status Register when all pending device operations have finished.
Query Syntax
*OPC?
The *OPC? query places an ASCII "1" in the output queue when all
pending device operations have completed. The interface hangs until this
query returns.
Return Format
<complete><NL>
<complete> ::= 1
See Also "Introduction to Common (*) Commands" on page 99
"*ESE (Standard Event Status Enable)" on page 102
"*ESR (Standard Event Status Register)" on page 104
"*CLS (Clear Status)" on page 101

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Related product manuals