Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 343
:MTESt:RMODe:WAVeforms
(see page 658)
Command Syntax
:MTESt:RMODe:WAVeforms <count>
<count> ::= number of waveforms in NR1 format
from 1 to 2,000,000,000
When the :MTESt:RMODe command is set to WAVeforms, the
:MTESt:RMODe:WAVeforms command sets the number of waveform
acquisitions that are mask tested.
Query Syntax
:MTESt:RMODe:WAVeforms?
The :MTESt:RMODe:WAVeforms? query returns the number of waveforms
currently set.
Return Format
<count><NL>
<count> ::= number of waveforms in NR1 format
from 1 to 2,000,000,000
See Also "Introduction to :MTESt Commands" on page 320
":MTESt:RMODe" on page 337