EasyManuals Logo

Agilent Technologies InfiniiVision 5000 Series Programmer's Guide

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #337 background imageLoading...
Page #337 background image
Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 337
:MTESt:RMODe
(see page 658)
Command Syntax
:MTESt:RMODe <rmode>
<rmode> ::= {FORever | SIGMa | TIME | WAVeforms}
The :MTESt:RMODe command specifies the termination conditions for the
mask test:
FORever — the mask test runs until it is turned off.
SIGMa — the mask test runs until the Sigma level is reached. This level
is set by the ":MTESt:RMODe:SIGMa" on page 341 command.
TIME — the mask test runs for a fixed amount of time. The amount of
time is set by the ":MTESt:RMODe:TIME" on page 342 command.
WAVeforms — the mask test runs until a fixed number of waveforms are
acquired. The number of waveforms is set by the
":MTESt:RMODe:WAVeforms" on page 343 command.
Query Syntax
:MTESt:RMODe?
The :MTESt:RMODe? query returns the currently set termination condition.
Return Format
<rmode><NL>
<rmode> ::= {FOR | SIGM | TIME | WAV}
See Also "Introduction to :MTESt Commands" on page 320
":MTESt:RMODe:SIGMa" on page 341
":MTESt:RMODe:TIME" on page 342
":MTESt:RMODe:WAVeforms" on page 343
Example Code "Example Code" on page 320

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals