:MEASure:STATistics
(see page 658)
Command Syntax
:MEASure:STATistics <type>
<type> ::= {{ON | 1} | CURRent | MINimum | MAXimum | MEAN | STDDev
| COUNt}
The :MEASure:STATistics command determines the type of information
returned by the :MEASure:RESults? query. ON means all the statistics are
on.
Query Syntax
:MEASure:STATistics?
The :MEASure:STATistics? query returns the current statistics mode.
Return Format
<type><NL>
<type> ::= {ON | CURR | MIN | MAX | MEAN | STDD | COUN}
See Also "Introduction to :MEASure Commands" on page 272
":MEASure:RESults" on page 291
":MEASure:STATistics:RESet" on page 301
":MEASure:STATistics:INCRement" on page 300
Example Code "Example Code" on page 291