170 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:ACQuire:POINts
(see page 658)
Query Syntax
:ACQuire:POINts?
The :ACQuire:POINts? query returns the number of data points that the
hardware will acquire from the input signal. The number of points
acquired is not directly controllable. To set the number of points to be
transferred from the oscilloscope, use the command :WAVeform:POINts. The
:WAVeform:POINts? query will return the number of points available to be
transferred from the oscilloscope.
Return Format
<points_argument><NL>
<points_argument> ::= an integer in NR1 format
See Also "Introduction to :ACQuire Commands" on page 163
":DIGitize" on page 132
":WAVeform:POINts" on page 512